- This test method covers the determination of the amount of carbon residue formed after evaporation and pyrolysis of petroleum materials under certain conditions and is intended to provide some indication of the relative coke forming tendency of such materials. The test results are equivalent to the Conradson carbon residue test.
- Glass Sample Vials : 2ml capacity, 12mm outside diameter by approximately 35mm height. Larger 4-dram glass sample vials, 15ml capacity (20.5 to 21mm outside diameter by 70 +1mm height) may be used for samples that are expected to yield residues <0.10% (m/m) so that a more appreciable mass difference can be determined. It should be noted the precision statements for the test method were determined using the 2ml capacity vials only (with sample residues between 0.3 and 26% (m/m) and that the precision associated with using the larger sample vials has not been determined.
- Coking Oven : With circular heating chamber approximately 85mm in diameter by 100mm deep, for top loading, with heating capability from 10 to 40°C/min rate to 500°C, with exhaust port 13mm inside diameter for nitrogen purge of oven chamber (inlet near top, exhaust at bottom center) with thermo couple sensor located in oven chamber next to but not touching sample vials, with lid capable of sealing out air and with removable condensate trap located at the oven chamber base.
- Sample Vial Holder : Cylindrical aluminum block, approximately 76mm in diameter by 16mm thick, with 12 evenly spaced holes (for vials) each 13mm in diameter by 13mm deep. The holes are arranged in a circular pattern approximately 3mm from the perimeter. The holder has legs 6mm long, with guides to center in oven chamber and an index mark on the side to use as position reference.
- Power supply : 230 +10% Volt AC, 1 Phase, 50 Hz.
| Equipment Dimensions | Shipping Dimensions | Net Weight | Shipping Weight |
| 60 W x 50 D x 65 H Cm | 75 W x 65 D x 80 H Cm | 55.0 Kg |
85.0 Kg |
| 5 |
|
0 |
| 4 |
|
0 |
| 3 |
|
0 |
| 2 |
|
0 |
| 1 |
|
0 |
Related Products
- Product Code: Qnix 4600
QNix® 4600 – The coating thickness gaugeThe QNix® 4600 is a first-class measuring device that can easily keep up with its test objects thanks to its high speed of over 100 measurements per minute. The handy and ergonomic design of the device makes it the ideal choice for vehicle evaluation and secures it a pole position in terms of performance and ease of use. The practical one-button operati..
NHG268 tri angle touch screen gloss meter features with 20, 60 and 85° angle, manufactured according to ISO2813 and GB/T 9754. It is also compatible with the standards of ASTM D523, ASTM D2457. NHG268 touchscreen glossmeter conforms to JJG696 first class gloss meter working requirement, corresponding with CIE 1931(2°) under CIE C light source. NHG268 touch screen gloss meter 2000gu can test..
NH310-High-End Brand of Domestic Colorimeter NH310 is a mainstream brand colorimeter introduced by 3nh which has synthesized the advantages of ten more traditional imported colorimeters and has experienced ten thousands tests and applied many innovative technologies. Accurate, Stable, Exquisite and Affordable! 8/b Viewing Geometry. Φ 8mm / Φ 4mm Measuring Aperature. (Φ8 mm is optional)..
100 in stock
NHG268 gloss meter features with 20, 60 and 85° angle, manufactured according to ISO2813 and GB/T 9754. It is also compatible with the standards of ASTM D523, ASTM D2457. NHG268 conforms to JJG696 first class gloss meter working requirement, corresponding with CIE 1931(2°) under CIE C light source. NHG268 gloss meter can test material with a high gloss (0-2000Gu), and universally applicable..






Reviews
There are no reviews yet.